Cell countingThis is a featured page

HEMO CYTOMETER & CELL COUNTING

A hemacytometer (also spelled hemocytometer) is an etched glass chamber with raised sides that will hold a quartz coverslip exactly 0.1 mm above the chamber floor. The counting chamber is etched in a total surface area of 9 mm2 (see Figure 1). Cell Culture - Tricksters
Figure 1. Dimensions of a hemacytometer. Calculation of concentration is based on the volume underneath the cover slip. One large square (see W in Figure 2) has a volume of 0.0001 ml (length x width x height; i.e., 0.1 cm x 0.1 cm x 0.01 cm). In both methods, the hemacytometer is filled by capillary action - place the pipette filled with a well-suspended mix of cells at the notch at the edge of the hemacytometer and then slowly expel some contents so that the fluidis drawn into the chamber by capillary action. Staining of cells often facilitates visualization and counting. Either mix cells with an equal volume of trypan blue [0.4% (w/v) tyrypan blue in PBS] to determine live/dead count (dead cells are blue) or kill cells with 10% formalin and then stain with trypan blue or other stain (to improve visualization of all cells. Here are two simple methods for counting cells based on the surface area of the hemacytometer used to determine cell count. Other counting schemes are accetable also. The choice of methods depends upon the cell concentration - the accuracy of the procedure depends upon the number of cells counted. When cell concentration is low, one should count more grids.Method A
Count the number of cells in the 4 outer squares (see the left panel of Figure 2). The cell concentration is calculated as follows: Cell concentration per milliliter = Total cell count in 4 squares x 2500 x dilution factorExample: If one counted 450 cells after diluting an aliquot of the cell suspension 1:10, the original cell concentration = 450 x 2500 x 10 = 11,250,000/ml Method B
Estimate cell concentration by counting 5 squares in the large middle square (see the right panel in Figure 2). The cell concentration is calculated as follows: Cell concentration per milliliter = Total cell count in 5 squares x 50,000 x dilution factor Example: If one counted 45 cells after diluting an aliquot of the cell suspension 1:10, the original cell concentration = 45 x 50,000 x 10 = 22,500,000/mlCell Culture - Tricksters
Figure 2. Counting procedure for Methods A (left panel) and B (right panel).


No user avatar
amitbalakrishnan
Latest page update: made by amitbalakrishnan , Dec 29 2006, 3:49 PM EST (about this update About This Update amitbalakrishnan Edited by amitbalakrishnan

405 words added
2 images added

view changes

- complete history)
More Info: links to this page
There are no threads for this page.  Be the first to start a new thread.

Related Content

  (what's this?Related ContentThanks to keyword tags, links to related pages and threads are added to the bottom of your pages. Up to 15 links are shown, determined by matching tags and by how recently the content was updated; keeping the most current at the top. Share your feedback on Wetpaint Central.)